FM Approved and CSA Certified as intrinsically safe with entity, for use in Class I, II, III, Division 1, Groups A, B, C, D, E, F, G; T-codes: FMRC = T3A, CSA = T4; hazardous locations. Nonincendive for use in Class I, II, III, Division 2, Groups A, B, C, D, F, G.
MAX. ENTITY PARAMETERS
Vmax = 30 V, Imax =175 mA, Ci = 0, Li = 0
ENCLOSURE
Impact-resistant glass filled polycarbonate body; impact resistant clear polycarbonate cover; NEMA & CSA 4X, IEC 529, IP67, ½" conduit hole provided at base. May be provided on back for panel mounting applications, call factory for details.
DISPLAY
Sharp, large 1.0" (25.4 mm) high LCD, ±1999
INPUT
4-20 mA
CALIBRATION
2 Step; non-interacting zero and span
CALIBRATION RANGE
4 mA input: -1000 to +1000; 20 mA between 20 and 2000 counts >4 mA display
ACCURACY
±0.1% of span, ±1 count
CONVERSION RATE
2.5 Conversions/second
MAXIMUM INPUT CURRENT
30 mADC
MAXIMUM VOLTAGE DROP
1 V at 20 mA
CONNECTIONS
Screw terminal block
OPERATING TEMPERATURE RANGE
-40 to 85°C (-40 to 185°F); -40 to 40°C (-40 to 104°F) for Canadian installations
WARRANTY
1 year parts and labor.
EXTENDED WARRANTY
May be extended an additional 12 months by returning the Product Registration Form within 2 months from date of purchase.