Wilkerson Instrument Co. Homepage A World of Signal Conditioning & Process Control Instrumentation Products


MM4001
RTD INPUT
TRANSMITTER

FEATURES
bulletProvides DC Output Proportional to an RTD Input
bullet3-Wire or 2-Wire, 10 ohms to 2000 ohms RTDs
bulletUnlimited* Choice of Input/Output Ranges
bulletLinearized for Platinum RTDs
bulletFully Isolated Output Available
bulletChoice of Power Options
bullet10 + 5 Year Warranty


DESCRIPTION

The MM4001 provides an output voltage or current proportional to an RTD input signal. It is useful in measuring temperatures with a variety of RTDs.

The module includes filtering and conditioning to reduce susceptibility to transients and noisy operations.

The module utilizes a single constant current source to excite the RTD. An accurate and stable lead wire compensation circuit uses the third lead of the 3-wire RTD to compensate for the RTD lead resistance.

TYPICAL APPLICATIONS

Temperature control, monitoring, and data acquisition.

MOUNTING

The module is designed to plug into a standard 8-pin relay socket. Part number MP008 is a molded plastic socket suitable for mounting on a flat surface or in a 2 3/4 inch wide PVC snap track (TRK48).  Use CLP-1 hold-down clip if needed. A Killark HK Series explosion-proof housing with dome and 8-pin socket is available (HKB-HK2D-8).  A DIN rail mounted socket (DMP008) is available for 35mm symmetrical DIN rail.

SPECIFICATIONS

RTD INPUT
3-wire or 2-wire
10 ohms to 2000 ohms

INPUT RANGE
select any range within RTD limit
  min span 25°F/14°C
  (100°F/55°C with 10 ohms RTD)

EXCITATION CURRENT
10 ohms
100 ohms
1000 ohms
2000 ohms
10 mA
5 mA
0.5 mA
0.2 mA

OUTPUT RANGE
Voltage
  select any range from  -10 V to +15V
  10 mA max load
    (min span 0.2 V)

Current
  select any range from 0 to 50 mA max
    (min span 1 mA)
  24 V compliance**
    18 V compliance for ISO option
    if full scale output >20 mA

OPEN SENSOR OUTPUT
> = full scale

ACCURACY
±0.1% of span

LINEARITY
(Pt RTD, output vs temp)
±0.05% of span
  temp > = 32°F/0°C

±0.15% of span
  temp < = 32°F/0°C

(others, output vs res)
±0.01% of span

COMMON MODE REJECTION
120 dB, DC to 60 Hz

ISOLATION (Optional)
Output/Input
  >500 megohms

Breakdown Voltage
  >1000 VAC rms

OUTPUT RIPPLE
<0.1% of span
  (peak to peak)

BREAKDOWN, P0WER/CIRCUITRY
>1500 VAC rms

OPERATING TEMPERATURE
 14°F to 140°F
-10°C to  60°C

TEMPERATURE STABILITY
±0.02% of span or 0.025 °C/°C
  whichever is greater

POWER
115 VAC ±10%, 50/60 Hz
  (2.5 W max)

230 VAC ±10%, 50/60 Hz
  (2.5 W max)

DC Power Option
  24 VDC
    (limits 21-32 VDC)
    (2.5 W max)
  Isolation
    DC power supply to input common:
      10 megohms

* Within specified range limits

** Compliance:  The sum of all voltage drops in the output loop cannot exceed 24V at rated current.


CONNECTIONS
Power AC L1 or DC +
Power AC L2 or DC -
RTD Lead Compensation
RTD -
RTD +
Output +
Output -
PINS
1
3
4
5
6
7
8